Operation Manual

For

NDJ-1 Rotary Viscometer

MADE  IN  CHINA

Thank you for using NDJ-1 Rotary Viscometer. NDJ-1 is easy to use, in order to install and operate accuray, and make full use of product features, we suggest you to read the manual before using.

· The company possesses the final interpretation power.

· The company reserves the power of amending technical rules without prior notice.

· The company reserves the power to amend the manual without further notice.

· May not be translated into other languages, reproduction prohibited without written permission. 

Cataloge

1General……………………………………………………

2Main technical data…………………………………………

3Principle………………………………………………

4Installation……………………………………………………

5Operation procedures………………………

6Precaution………………………………………………

Packing List ……………………………………………………

1. General

NDJ-1 rotary viscometer is a new instrument used for determining the liquid viscose capacity and the absolute viscosity. NDJ-1 has been widely used to determine and measure the liquid viscosity in many applications such as grease, painting, plastic, pharmacy and adhesives. It is a precision instrument to monitor and control the stable quality of products in the manufacturing.

2. Main technical data

1) Measurement range: 10-100.000 mPa·s 

2) Rotor specification: 0, 1, 2, 3, 4, 5, five kind rotor (0# rotor could measure the viscometer from low viscometer to 0.1 mPa·s)

3) Measurement error: ±5% (Newton liquid)

4) Power supply: Voltage--220±22V, Frequency50 Hz±0.5Hz

5) Dimensions: 300 mm ×300 mm × 450 mm

6) Net weight: 1.5 Kg (the base not included)

3. Principle

1) Figure 1Synchronous motor rotates with stable rate, connect scale disk, through hairspring and rotating shaft, drives rotors to rotate. The rotors will subject to a torque moment proportional to liquid viscosity because of the liquid viscose hysteresis. The torque moment will be measured by the sensors and processed into the viscosity and shown on the display.

2) Rate can be changed by using gear system and clutch, operated by rotary knob, four optional velocity.

3) According to different equipments, different rotors (Five rotors: NO.0-NO.4) will be attached, it can be select with the velocity according to the viscosity of the liquid.

4) The fixed control gear of pointer are used for accurate reading. When the rate is so fast (30 rpm, 60 rpm) that you can not get the reading while it rotated, press pointer joystickfix the pointer in order to get the reading.

5) The rotor bracket is used for the protection of rotors and the stability of the measurement, a more reliable measuring result can be made by using the protection bracket. 

6) It can be used as a portable instrument, assigned fixed bracket and elevating system. Generally, it should be fixed when determine small amount or the temperature is fixed in the laboratory.

7) The instrument has gears and damping elevate mechanism. It is stable for instrument lifting.

8) The whole set of instrument has good quakeproof packing to ensure the precision of instrument.

Installation

1. Take out the bracket, package, stanchion and so on from the packaging.

2. Screw the column to the threaded hole behind the bracket, and put the tooth profile surface in the column facing front the bracket. Screw the hex nut tight by wrench to prevent the column running.

3. Rotational move the lifting chuck knob to check the flexibility and self-looking of lifting chuck. Found the lifting chuck too loosen and too tight phenomenon, can be used hex nut wrench to adjust the tight screw. So that it can lift from top to bottom. And it is suitable for tight to prevent the situation of automatically falling after installing the viscometer.

4. Take out viscometer form package and mount it on the lifting collettight it by handle setscrew to avoid loosen(as level as possible), link the power supply, take away the rubber band on the pointer joystick, screw the nut of protecting-cap loosen under the instrument, take away the protecting -cap.

5. Adjust the adjustment screw and put the rotor into the liquid to be measured till the level mark on the rotor reach the liquid surface.

Operation procedures

1. Prepare the liquid to be measured and put it into a glass beaker or a right angle container with the diameter not smaller than 70 mm, height not smaller than 130mm, take care of the liquid temperature.

2. Mount the protection bracket on instrument, turning right for mounting, turning left for removing.

3. Screw the selected rotor into connecting rod, turning right for mounting, turning left for removing , Adjust the lifting screw and put the rotor into the liquid to be measured till the level mark on the rotor reach the liquid surface. Connect power supply, turn on the equipment, screw the speed knob, select velocity, relax pointer joystick, reading data can be displayed (about 20~30 seconds) when the pointer becomes stable. Press the pointer joystick after pointer becoming stable, cut power supply, then get the reading data. If the pointer isnt in reading window after motor stop, you can press pointer joystick unceasingly, open or close the motor time after time, after practice, you can operate it familiarly.

4. When the reading data is too high or too low, the rotor or the rate can be changed, make sure the reading is between 20 and 80.

5. Choice of range, coefficient, rotor and rate:

5.1 Estimate approximay the viscose range, then select the rotor and the velocity based on the range table given below:

For example, the viscosity of liquid is about 3000 mPa·s the settings should be as following:  

rotor 2 # with velocity of 6 rpm,

rotor 3 # with velocity 0f 30 rpm

5.2 If the viscosity of liquid can not be estimated, it should take a high value for real measurement, the rotor selection should be made from smaller to larger with their number from higher to lower. In general, high viscosity should use smaller rotor with slower velocity, lower viscosity use lager rotor with faster velocity.

5.3 Coefficient: the reading must be multiplied the specific coefficient in the coefficient table in order to get the absolute viscosity.

   That is: 

            absolute viscosity

            coefficient

            reading ( deflection angle)

   5.4 Amendment of frequency error: when the power frequency is inaccurate, you can amendment it according to the formula below:

      Actual viscosity= instruction viscosity*nominal frequency/ actual frequency

Velocity

Range

Rotor

60

30

12

6

1

100

200

500

1000

2

500

1000

2500

5000

3

2000

4000

10000

20000

4

10000

20000

50000

100000

   5.5 

5.6 Coefficient table: 

        Rpm

Rotor

60

30

12

6

1

1

2

5

10

2

5

10

25

50

3

20

40

100

200

4

100

200

500

1000

 

Precaution

1. This instrument is limited to the room temperatures

2. This instrument should be used under the designed voltage and frequency and their allowable error ranges, or incorrect results could be resulted.

3. Using the bracket fix the instrument in order to determine accuray. Handheld operation: keep the instrument stable and level.

4. Care should be taken for mounting or removing rotors, slightly and lift the connecting screw bolt to avoid a transverse force acting on rotor to cause it bending.

5. The instrument mounted with rotor should not be revolved.

6. Don’t run the motor when the pointer joystick is not pressed.

7. Keep cleaned on screws and connecting points between rotor and connecting bolt rod, or a unstable rotation could be caused in the real measurement.

8. Use hand to hold the instrument when lifting up or moving down to avoid dropping it.

9. After completing measurement each time, the rotor should be fully cleaned (rotor should be removed from instrument for cleaning), then place it on the protection bracket.

10. Unauthorized removing or replacing the instrument part, and applying lubricates are not allowable.

11. When moving or shipping instrument, put on the yellow caver cap and lift the connecting bolt rod and screw the bolt on the cap tightly.

12. Suspension, emulsion or polymer and other high viscosity liquids are non-Newton liquids, their viscosity will change with shear velocity and time, it is normal for their measured inconsistent results under the selected rotor and velocity and time, and it is not resulted from the instrument problems (in general, the rotor and velocity and time should be specified for non-Newton liquids).  

13. The cautions should be taken for followings to obtain a good measuring reault:

(1) Accuray control the temperature of liquid to be measured;

(2) Make temperature of rotor and protection bracket same as liquid;

(3) Ensure the liquid homogeneity and without bubbles;

(4) Put the rotor located on the center of the liquid container in real measurement;

(5) Remove bubbles adhered on the rotor when put it into liquid;

(6) When change the rotor or rate, keep between 30~90 as possible.

(7) Use rotor protection bracket for measurement;

(8) Ensure rotor and connecting screw rod cleaned;

(9) Amend according to the amending formula when power frequency is not exact.

(10) When pointer is stable, reading with the same angle of view.

(11) Adjust level bubble exactly, move the instrument slowly, keep the instrument stable.

(12) Strictly follow the operation instruction for measurement.

Packing List

1) NDJ-1 Digital Rotary Viscometer           1 Qt.

2) Rotors, 1#, 2#, 3# and 4#                   1 Qt of each

3) Protection bracket                       1 Qt.

4) Bracket                                1 Qt.

5) Adjusting screw                         3 Qt.

6) Operation manual                        1 Qt.

7) License                                1 Qt.

8) Warranty card                           1 Qt.

版權所有©2025 上海平軒科學儀器有限公司   滬ICP備10019381號-3   sitemap   管理登陸   技術支持:化工儀器網   平軒科儀產品供應上海、杭州、北京、天津、深圳、廣州等全國各地
主站蜘蛛池模板: 14萝自慰专用网站 | 久久久久久久黄色 | 黑人操日本 | 曰欧一片内射vα在线影院 国产做爰视频免费播放 | 亚洲高清在线看 | 亚洲综合无码精品一区二区三区 | 久久艹精品视频 | se在线观看| 尤物视频在线免费观看 | 黑人中文字幕一区二区三区 | 久久人人爽人人爽人人片dvd | 欧美三级不卡在线观看 | 在线不卡免费视频 | 美女裸体色黄污视频网站 | 亚洲伊人情人综合网站 | 好吊妞这里只有精品 | 亚洲欧美偷拍视频 | 亚洲和欧洲一码二码区别7777 | 另类亚洲小说图片综合区 | 亚洲成a人片77777群色 | 国产在线精品观看免费观看 | 欧美视频网站 | 少妇裸交aa大片 | 饥渴的熟妇张开腿呻吟视频 | 国产 av 仑乱内谢 | 亚洲综合在线中文字幕 | 国产女人毛片 | 综合色区国产亚洲另类 | 午夜影院在线观看 | 精品久久久久成人码免费动漫 | 91av在线免费观看 | 伊人无码一区二区三区 | 精品亚洲国产成人蜜臀优播av | 夜夜嗨网站 | 性人久久网av| 看黄色网址 | 狠狠色老熟妇老熟女 | 亚洲一区二区精品 | 亚洲成年人在线 | 免费一区二区三区四区 | 中文字幕无码精品亚洲资源网 | 国产无遮挡又黄又大又不要vip | 国产精品va无码一区二区 | 天堂中文在线8 | 欧美黄在线观看 | 乱码精品一区二区三区 | 在线小视频国产 | 强插女教师av在线 | 久久久亚洲成人 | 六月丁香综合 | a√在线视频 | 正在播放国产精品 | 日本午夜免费啪视频在线 | 久久精品无码精品免费专区 | 久久综合a∨色老头免费观看 | 国产三级三级三级三级三级 | 亚洲高清有码中文字 | 制服丝袜亚洲中文欧美在线 | 在线免费看污片 | 久久亚洲中文字幕精品一区 | 中文字幕一区二区三区精品 | 又爽又黄又高潮的免费视频 | 国产午夜片无码区在线观看爱情网 | 老司机在线精品视频网站 | 国产a黄 | 精品午夜福利无人区乱码一区 | 亚洲大胆人体 | 日日操夜夜干 | 自拍偷拍亚洲图片 | 国产天美传媒性色av出轨 | 久久精品九九亚洲精品天堂 | 天天有av | 无码人妻精品一区二区三区久久 | 亚洲色大成网站www永久在线观看 | 亚洲第一av无码专区 | 天堂在线最新版www中文 | 91黄漫 | 色噜噜狠狠综曰曰曰 | 希岛爱理av一区二区三区 | 亚洲精品一区 | 99v久久综合狠狠综合久久 | xx久久 | 亚洲日批视频 | 中文字幕人成乱码熟女香港 | 久久伊人精品一区二区三区 | 好男人在在线社区www在线影院 | 欧美一区二区三区在线看 | 亚洲国产欧洲 | 亚洲成人av综合 | 国产亚洲成av人在线观看导航 | 亚洲婷婷五月综合狠狠app | 中文字幕亚洲欧美 | 精品欧洲av无码一区二区 | 久久久久久国产精品无码下载 | 欧美在线观看一区 | 日韩激情四射 | 在线 | 麻豆国产传媒61国产免费 | 青青视频在线免费观看 | 麻豆国产va免费精品高清在线 |